A part of MilliporeSigma

KHGLS1TBB1 | Opticap® Sterile Multilayer XLT10 Durapore® 0.45/0.22 µm 1 in. HB/HB

1  Double Easy-Open bag
Retrieving price...
Price could not be retrieved
Minimum Quantity needs to be mulitiple of
Upon Order Completion More Information
You Saved ()
Request Pricing
Limited AvailabilityLimited Availability
Limited Quantities Available
    Remaining : Will advise
      Remaining : Will advise
      Will advise
      Contact Customer Service


      Contact Customer Service

      Click To Print This Page


      Replacement Information

      Key Specifications Table

      Catalogue NumberKHGLS1TBB1
      Trade Name
      • Opticap®
      DescriptionOpticap® Sterile Multilayer XLT10 Durapore® 0.45/0.22 µm 1 in. HB/HB
      Product Information
      Device ConfigurationCapsule
      Connections, Inlet/Outlet25 mm (1 in.) Hose Barb
      Connections, Vent/Drain1/4 in. Hose Barb with double O-ring Seal
      Housing MaterialPolypropylene (PP)
      Maximum Differential Pressure, bar (psid)Forward: 5.5 bar (80 psid) @ 25 °C; 1.05 bar (15 psid) @ 80 °C. Reverse: 3.5 bar (50 psid) @ 25 °C, intermittent
      Maximum Inlet Pressure, bar (psi)5.5 bar (80 psi) @ 23 °C; 2.75 bar (40 psi) @ 60 °C; 1.0 bar (15 psi) @ 80 °C
      European Pressure Equipment DirectiveEMD Millipore Corporation® certifies that this product complies with the European Pressure Equipment Directive, 97/23/EC of 29 May 1997. This product has been classified under Article 3 § 3 of the Pressure Vessel Directive. It has been designed and manufactured in accordance with sound engineering practice to ensure safe use. In compliance with Article 3 § 3 of this Pressure Vessel Directive, this product does not bear the CE mark.
      Good Manufacturing PracticesThese products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.
      T-line/Gauge PortT-line
      Biological Information
      Bacterial EndotoxinsAqueous extraction contains ≤0.25 EU/mL as determined by the Limulus Amebocyte Lysate (LAL) Test
      MediaMultilayer Durapore®
      Sterilization3 autoclave cycles of 60 min @ 126 °C; not in-line steam sterilizable
      Bacterial RetentionSamples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
      Physicochemical Information
      Pore Size0.45 / 0.22 µm
      Inlet ConnectionHose Barb
      Outlet ConnectionHose Barb
      Air Diffusion at 23 °C≤15.0 mL/min @ 2.75 bar (40 psig) in water
      Bubble Point at 23 °C≥3450 mbar (50 psig) air with water
      Indirect Food AdditiveAll component materials meet the FDA Indirect Food Additive requirements cited in 21 CFR 177-182.
      Maximum Inlet Pressure (psig)80 psig
      Maximum Inlet Temperature25 °C
      Non-Fiber ReleasingThis product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).
      TOC/ConductivityAfter sterilization and a controlled water flush, samples exhibited less than 500 ppb TOC per USP <643> and less than 1.3 µS/cm per USP <645> at 25 ºC.
      USP Bacterial EndotoxinsA sample aqueous extraction contains less than 0.25 EU/mL as determined by the Limulus Amebocyte Lysate (LAL) test.
      Cartridge Nominal Length10 in. (25 cm)
      Length37.6 cm (14.8 in.)
      Diameter10.7 cm (4.2 in.)
      Filtration Area0.55 m²
      Inlet to Outlet Width19.8 cm (7.8 in.)
      Device Size10 in.
      Inlet Connection Diameter1 in.
      Outlet Connection Diameter1 in.
      Materials Information
      • Polyvinylidene Fluoride (PVDF)
      Device Material
      • Polypropylene
      Seal MaterialSilicone (SI)
      Support MaterialPolypropylene
      Vent Cap MaterialPolypropylene (PP)
      Toxicological Information
      Safety Information according to GHS
      Safety Information
      Product Usage Statements
      Storage and Shipping Information
      Packaging Information
      Material Size1
      Material PackageDouble Easy-Open bag
      Transport Information
      Supplemental Information
      Oxidizable SubstancesWill meet the USP Oxidizable Substances Test requirements after a water flush of ≥1.5 L