CBA086 Glucose Assay Kit

CBA086
Price could not be retrieved
Minimum Quantity needs to be mulitiple of
Upon Order Completion More Information
You Saved ()
 
Request Pricing
Limited AvailabilityLimited Availability
Stocked 
Discontinued
Limited Quantities Available
Available
    Remaining : Will advise
      Remaining : Will advise
      Will advise
      Contact Customer Service
      Contact Customer Service
      View Pricing & Availability

      Overview

      Replacement Information

      Key Specifications Table

      Detection Methods
      Colorimetric or Fluorimetric

      Pricing & Availability

      Catalog Number AvailabilityPackaging Qty/Pack Price Quantity
      CBA086-1KIT
      Retrieving availability...
      Limited AvailabilityLimited Availability
      Stocked 
      Discontinued
      Limited Quantities Available
      Available
        Remaining : Will advise
          Remaining : Will advise
          Will advise
          Contact Customer Service
          Contact Customer Service

          Glass bottle 1 kit
          Retrieving price...
          Price could not be retrieved
          Minimum Quantity needs to be mulitiple of
          Upon Order Completion More Information
          You Saved ()
           
          Request Pricing
          Description
          OverviewThe Glucose Assay kit provides direct measurement of glucose in various biological samples (e.g. serum, plasma, body fluids). Glucose enzyme mix oxidizes glucose to generate a product which reacts with a dye to produce color and fluorescence. The generated color or fluorescence is proportional to the amount of glucose in the sample.
          Catalogue NumberCBA086
          Brand Family Calbiochem®
          Application Data
          References
          ReferencesLee, Y.S. et al. 2007.Diabetes 56, 1671.
          Product Information
          Detection methodColorimetric or Fluorimetric
          Form100 tests
          Format96-well plate
          Kit containsGlucose Assay Buffer, Glucose Probe, Glucose Enzyme Mix, and Glucose Standard, and a user protocol.
          Applications
          Biological Information
          Assay range1-200 µM
          Physicochemical Information
          Dimensions
          Materials Information
          Toxicological Information
          Safety Information according to GHS
          Safety Information
          R PhraseR: 21/22-36/38

          Harmful in contact with skin and if swallowed.
          Irritating to eyes and skin.
          S PhraseS: 24/25-26-36-45

          Avoid contact with skin and eyes.
          In case of contact with eyes, rinse immediately with plenty of water and seek medical advice.
          Wear suitable protective clothing.
          In case of accident or if you feel unwell, seek medical advice immediately (show the label where possible).
          Product Usage Statements
          Intended useThe Glucose Assay Kit provides direct measurement of glucose in various biological samples (e.g., serum, plasma, body fluid, food, growth medium, etc.). The method is rapid, simple, sensitive, and suitable for high throughput. The assay is also suitable for monitoring glucose levels during fermentation and glucose feeding in protein expression processes.
          Storage and Shipping Information
          Ship Code Dry Ice Only
          Toxicity Multiple Toxicity Values, refer to MSDS
          Storage -20°C
          Storage ConditionsUpon arrival store the entire contents of the kit at (-20°C). Protect from light. Note: Allow reagents warm to room temperature and briefly centrifuge vials before opening.
          Do not freeze Ok to freeze
          Mercury prohibited To comply with ban of sale of mercury-added products required by The Interstate Mercury Education and Reduction Clearinghouse (IMERC), this product is prohibited to be sold in the following US states: Rhode Island and Connecticut.
          Packaging Information
          Transport Information
          Supplemental Information
          Kit containsGlucose Assay Buffer, Glucose Probe, Glucose Enzyme Mix, and Glucose Standard, and a user protocol.
          Specifications

          Documentation

          Glucose Assay Kit SDS

          Title

          Safety Data Sheet (SDS) 

          Glucose Assay Kit Certificates of Analysis

          TitleLot Number
          CBA086

          References

          Reference overview
          Lee, Y.S. et al. 2007.Diabetes 56, 1671.
          User Protocol

          Revision17-September-2012 JSW
          Form100 tests
          Format96-well plate
          Detection methodColorimetric or Fluorimetric
          StorageUpon arrival store the entire contents of the kit at (-20°C). Protect from light. Note: Allow reagents warm to room temperature and briefly centrifuge vials before opening.
          Intended useThe Glucose Assay Kit provides direct measurement of glucose in various biological samples (e.g., serum, plasma, body fluid, food, growth medium, etc.). The method is rapid, simple, sensitive, and suitable for high throughput. The assay is also suitable for monitoring glucose levels during fermentation and glucose feeding in protein expression processes.
          BackgroundGlucose (C6H12O6; FW: 180.16) is a very important fuel source for generating universal energy molecules ATP. Glucose level is a key diagnostic parameter for many metabolic disorders. Measurement of glucose can be very important in both research and drug discovery processes.
          Principles of the assayThe Glucose Assay Kit utilizes the Glucose Enzyme Mix to specifically oxidize glucose to generate a product that reacts with a dye to generate color (&lambda = 570 nm) and fluorescence (Ex/Em = 535/587 nm). The generated color and fluorescence is proportional to the glucose amount.
          Materials provided• Glucose Assay Buffer (Kit Component No. JA9439-25ML): 1 bottle, 25 ml
          • Glucose Probe (in DMSO, anhydrous) (Kit Component No. JA9440-1EA): 1 vial
          • Glucose Enzyme Mix (Kit Component No. JA9442-1EA): 1 vial
          • Glucose Standard (100 nmol/µl) (Kit Component No. JA9443-0.1ML): 1 vial, 100 µl
          PreparationPrepare test samples in a total volume of 50 µl/well with Glucose Assay Buffer. Serum can be directly diluted in Glucose Assay Buffer; typically 0.5-2 µl/assay is sufficient (normal serum contains ~5 nmol glucose/µl). We suggest testing several doses of your sample to make sure the readings are within the standard curve range.
          Reagent preparation• Glucose Enzyme Mix: Dissolve the Glucose Enzyme Mix in 220 µl Glucose Assay Buffer. Aliquot and store at -20°C. Use within two months.
          • Glucose Probe: Supplied ready to use. Warm to room temperature to thaw the DMSO.
          • Standard Curve: For colorimetric assay, dilute the Glucose Standard to 1 nmol/ml by adding 10 µl of the Glucose Standard to 990 µl of Glucose Assay Buffer, mix well. Add 0, 2, 4, 6, 8, 10 µl into each well individually. Adjust volume to 50 µl/well with Glucose Assay Buffer to generate 0, 2, 4, 6, 8, 10 nmol/well of Glucose Standard. For fluorometric assay, dilute the Glucose Standard solution to 0.1 nmol/µl by adding 10 µl of the Glucose Standard to 990 µl of Glucose Assay Buffer, mix well. Then take 20 µl into 180 µl of Glucose Assay Buffer. Mix well. Add 0, 2, 4, 6, 8, 10 µl into each well individually. Adjust volume to 50 µl/well with Glucose Assay Buffer to generate 0, 0.2, 0.4, 0.6, 0.8, 1.0 nmol/well of the Glucose Standard.
          Detailed protocol1. Prepare reagent mix as indicated below; prepare a volume that is sufficient for the number of assays to be performed; 50 µl per well is required. Each 50 µl of reaction mix should contain the following components:
          • 46 µl Glucose Assay Buffer
          • 2 µl Glucose Probe
          • 2 µl Glucose Enzyme Mix
          Mix well.
          2. Add 50 µl Reaction Mix to each well containing a diluted Glucose Standard or test sample(s). Mix well.
          3. Incubate the reaction for 30 min at 37°C, protect from light.
          4. Measure the absorbance at 570 nm for a colorimetric assay or at excitation/emission = 535/590 nm for a fluorometric assay using an appropriate microplate reader
          5. Correct for background by subtracting the value derived from the 'zero' glucose control from all Standard and sample readings. Note: The background reading can be significant and MUST be subtracted from Standard and sample readings. Glucose concentrations of the test samples can then be calculated based on the standard curve generated.
          Standard curve

          Figure 1: Standard Curve

          Assay Range1-200 µM
          Registered TrademarksCalbiochem® is a registered trademark of , KGaA, Darmstadt.
          InteractivePathways™ is a trademark of , KGaA, Darmstadt.